CLEANING, MAINTENANCE, OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
OVERHAUL
Shock waves are generated mechanically. Due to the effects of friction, the handpiece components are continuously exposed to
mechanical stress, which will cause minor wear.
Note: The Falcon
*
handpiece should be overhauled about every 1,000,000 shocks. This can be done by the user of the device. All that is required is
the overhaul kit, which includes all required wear parts.
Note: The sealing rings, the projectile and the guide tube must always be replaced each time the handpiece is overhauled. Observe the O-ring Guide
when selecting the sealing rings to be used. It is contained in the overhaul kit.
OVERHAULING THE HANDPIECE
• Flammable and volatile cleaning and disinfecting agents may generate an explosive atmosphere.
• Disconnect the handpiece from the control unit before starting any cleaning or maintenance work.
• Place the handpiece on a dry, clean and dust-free surface.
*Falcon is a registered trade mark of Storz medical AG
Attention!
An open end spanner has to be used to release or to assemble handpiece shaft when overhauling
the handpiece.