CLEANING, MAINTENANCE, OVERHAUL
EN
INTELECT
®
RPW 2 USER MANUAL
CLEANING THE HANDPIECE
Note: After cleaning, the handpiece must be dry before it can be
reassembled. For that reason, schedule sufficient time for the drying
of the handpiece and its components.
1. Disconnect the handpiece from the control device.
2. Unscrew the shock transmitter screw cap respectively
the clutch for the spine and fasciae shock transmitters from
the handpiece.
3. Clean the coupling gel off the handpiece using a cleaning
agent that is suitable for surface cleaning.
• Disinfect the handpiece with an alcohol-based
disinfectant that is suitable for surface cleaning
4. Unscrew the shaft from the handpiece and pull it out of
the handpiece handle (5).
• Use the supplied open-end spanner (4) for this
purpose.
5. Clean the guide tube with a brush in order to ensure
perfect projectile movement.
The assembly of the handpiece is carried out in reverse
order.
Note: When mounting the shaft of the handpiece it is necessary to
tighten it using the supplied open-end spanner The shaft may no
longer be solved manually.