Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
8 ULTRASOUND CALIBRATION
8.1 GENERAL
A. Tools and Equipment Required
Vectra Genisys, Intelect Legend XT and the
Intelect Vet Combination Therapy System
and all Vectra Genisys, Intelect Legend XT
and the Intelect Vet Ultrasound Applicators
associated with the System being serviced.
Ohmic Instruments UPM DT 10 or DT 100
Ultrasound Power Meter, set to “watts”.
Degassed Water.
Refer to page 37
for
Degassed Water Recipes.
B. Ultrasound Applicator Calibration
Procedures
Enter the Technical Service Screen of the
Therapy System by pressing the Stop,
Pause, and Start Buttons simultaneously.
See
Figure 8.1.
NOTE:
To access the Technical Service Screen of
Therapy Systems with Version 2.0 or above
software, simultaneously press the two
buttons located at the upper right of the
screen.
See inset at Figure 8.1
.
Set up Power Meter per Ohmic User Manual.
Position the Ultrasound Applicator in the
Power Meter.
Press the Ultrasound Calibration Button.
See
Figure 8.2.
Press the Head Size Button until the size
applicator being calibrated is displayed.
See
Figure 8.3.
Press the Start Button,
refer to Figure 8.3.
Follow the instructions displayed on the
Therapy System.
Repeat this procedure for each Ultrasound
Applicator associated with the Therapy
1.
2.
3.
1.
2.
3.
4.
5.
6.
FIGURE 8.1
FIGURE 8.2
FIGURE 8.3
ULTRASOUND CALIBRATION
BUTTON
HEAD SIZE BUTTON
START
BUTTON
THERAPY SYSTEMS WITH
SOFTWARE VERSION 2.0 OR
ABOVE
USE ONLY DEGASSED WATER IN POWER METER FOR
CALIBRATING ULTRASOUND APPLICATORS.
USE OF OTHER TYPES OF WATER WILL CAUSE FALSE
READINGS AND BAD TEST RESULTS.
SEE PAGE 37
FOR DEGASSED WATER RECIPES.
USE OF OTHER BRANDS OR TYPES OF TOOLS,
EQUIPMENT, FIXTURES, MATERIALS, AND SUPPLIES
OTHER THAN THOSE SPECIFICALLY LISTED IN “A. Tools
and Equipment Required” ABOVE WILL GIVE BAD
TEST AND CALIBRATION RESULTS.
IF PROPER EQUIPMENT IS NOT AVAILABLE OR
CAN NOT BE OBTAINED, SEND THE ULTRASOUND
APPLICATORS TO THE FACTORY FOR CALIBRATION.