19
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
4 SPECIFICATIONS
A. IFC (Interferential) Traditional (4 Pole)-
Figure 4.2 (Vectra Genisys, Intelect Legend
XT, and Intelect Vet)
Output Mode . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity . . . . . . . . . . . . . . . . 0-100 mA(CC)
0-100
V
(CV)
Carrier Frequency . . . .2,500, 4,000, and 5,000 Hz
Beat Frequency (Sweep Off ) . . . . . . . . . .1-200 Hz
Sweep Time (Fixed) . . . . . . . . . . . . . . . . 15 seconds
Sweep Low Beat Frequency . . . . . . . . . .1-199 Hz
Sweep High Beat Frequency . . . . . . . . . .2-200 Hz
Vector Scan . . . . . . . Off, Manual, 40%, and 100%
Treatment Time . . . . . . . . . . . . . . . . . . 1-60 Minutes
Mode Selection . . . . . . . . . . . . . . . . . . . . . CC or CV*
(Intelect Vet Mode Selection Fixed- CV only)
Available on Channels . . . . . .1 & 2, 3 & 4 Option
B. TENS- Asymmetrical Biphasic- Figure 4.3
(Vectra Genisys)
Output Mode . . . . . . . . . . . . . . . . . . . . . . . Electrodes
Output Intensity . . . . . . . . . . . . . . . . 0-110 mA (CC)
0-110 V (CV)
Phase Duration . . . . . . . . . . . . . . . . . .20-1,000 μsec
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-250 Hz
Mode Selection . . . . . . . . . . . . . . . . . . . . . CC or CV*
Burst Frequency . . . . . . . . . . . . . . . . . . . . . . 0-10 bps
Frequency Modulation . . . . . . . . . . . . . . .0-250 Hz
Amplitude Modulation . . . . . . . . . .Off, 40%, 60%,
80%, and 100%
Cycle Time . . . . . . . . . . . . . . . . . . . .4/4, 4/8, 7/7, 5/5,
4/12, 10/10, 10/20, 10/30, 10/50
Treatment Time . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channels . . . . . .1 & 2, 3 & 4 Option
The specifications found in this section provide
the necessary waveform specifications to aid in
troubleshooting. A waveform graphic from an
oscilloscope is also provided for clarification.
Refer to this section when performing
troubleshooting, replacement, and repair of the
Therapy System, Modules, and Accessories.
NOTE:
All waveforms, except High Voltage Pulsed Current
(HVPC), of the Vectra Genisys, Intelect Legend XT
and Intelect Vet Therapy Systems have a 200 mA
current limit.
VMS™, VMS™ Burst, and all TENS waveform output
intensities are measured, specified, and listed to
peak, not peak to peak.
All waveforms are available on all channels.
FIGURE 4.2
Stimulus delivered by the TENS waveforms of this device, in certain
configurations, will deliver a charge of 25 microcoulombs (μC) or
greater per pulse and may be sufficient to cause electrocution.
Electrical current of this magnitude must not flow through the
thorax because it may cause a cardiac arrhythmia.
FIGURE 4.3
*CC= Constant Current
CV= Constant Voltage
4.2 ELECTROTHERAPY WAVEFORM SPECIFICATIONS