Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
5 TROUBLESHOOTING
129
Warning
sEMG Data Card full.
A.
sEMG Data Card faulty. Insert a known good sEMG Data Card. If problem continues,
contact dealer or factory for service.
130
Warning
Another treatment is running while attempting to set up and
perform a Laser Therapy treatment.
A.
Allow existing treatment to complete before starting Laser Therapy.
B.
If no other treatment is running, reset Therapy System by turning main power switch
Off and On.
131
Warning
Treatment Room Door Lockout is breached.
A.
Make certain Treatment Room Door is completely closed.
B.
Make certain the Lockout cable is connected to the system.
C.
Replace Lockout to System cable with a known good cable.
D.
Contact department responsible for installation of the Treatment Room Door Lockout
mechanism for maintenance or repair.
E.
If problem continues, send Laser Module to factory for Service.
132
Warning
Attempted to start a laser treatment but noLaser Applicator is
plugged in.
A.
Connect desired Laser Applicator to the system.
B.
If Applicator is connected, reset Therapy System by turning main power switch Off and
On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for Service.
133
Warning
Laser Applicator became unplugged while performing a laser
treatment.
A.
Connect desired Laser Applicator to the system.
B.
If Laser Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for Service.
134
Warning
Entered incorrect laser PIN.
A.
Enter correct Laser PIN number.
B.
If problem continues, send Laser Module to factory for Service.
Code
Number
Type
Message
Probable Cause
Possible Remedies
118
Warning
Attempting to save additional User Protocols or Sequences after
system memory has reached the maximum allowed (200).
Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy
System User Manual for instructions.
119
120
121
122
Warning
Warning
Warning
Warning
1.
Attempted to read a treatment from Patient Data Card that is
not a valid treatment for the system
2.
Attempted to use a Non-Patient Data Card.
3.
No Patient Data Card inserted into system port.
4.
Unknown type of smart card inserted into system.
A.
Use a Patient Data Card with proper treatment data for the system.
B.
Properly insert a Patient Data Card.
C.
Insert a known good Patient Data Card.
D.
If problem persists, insert a known good Patient Data Card. If problem continues, contact
dealer or Factory for service.
123
Warning
Patient Data Card is full.
Erase Patient Data Card. Refer to Therapy System User Manual for instructions.
124
Warning
Patient Treatment Data already saved.
A.
Cannot save same data again on Patient Data Card.
B.
Use a new Patient Data Card to resave data.
C.
Erase Patient Data Card and resave treatment data.
125
Warning
Multimedia Card (MMC) not in system port.
A.
Properly insert the MMC card into the system port.
B.
Insert a known good MMC Card. If problem continues, contact dealer or Chattanooga
Group for Service.
126
Warning
No valid channels are available for attempted treatment.
A.
Complete existing treatment before attempting to start another.
B.
Reset Therapy System by turning main power switch Off and On.
127
128
Warning
Warning
1.
No sEMG Channels are available for treatment.
2.
No sEMG Module installed or detected by system.
A.
Wait until current treatment is complete.
B.
Reset Therapy System by turning main power switch Off and On.
C.
Make certain sEMG Module is properly installed. Refer to sEMG Module User Manual for
installation instructions.
D.
Replace sEMG Module with known good sEMG Module.
E.
If problem continues, contact dealer or factory for service.
5.1 THERAPY SYSTEM ERROR MESSAGES (CONTINUED)