63
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
6 REMOVAL/REPLACEMENT
A. Tools and Equipment Required
#1 Phillips Screwdriver
Flat Blade Screwdriver
B. sEMG Installation
NOTE:
If installing or replacing an sEMG Module
on a Therapy System with a Channel 3/4
Electrotherapy Module, Battery Module, or
Laser Module already installed, and the sEMG
functions are desired for Channels 1 and 2, it
will be necessary to remove the module from
the system.
Refer to 6.1, B
for instructions.
If replacing or installing an sEMG Module
to a Therapy System with a Channel 3/4
Electrotherapy Module, and it is desired that
sEMG function be available to Channels 3 and 4,
it will not be necessary to remove the module
as it will be installed to the Module.
The sEMG Module can only be installed on the
Therapy System and Channel 3/4 Electrotherapy
Module. This allows a maximum of two sEMG
modules on a four channel Electrotherapy or
Combination Therapy System. An sEMG Module
cannot be installed on the NiMH Battery or
Laser Module.
Only two channels of sEMG can be used at
any given time even if the Therapy System is
equipped with two sEMG Modules.
Disconnect Power Cord from the power
source and turn system over on its face.
NOTE:
When turning over Therapy System onto its
face, place a clean, soft cloth under the lens
to prevent scratching or lens damage.
1.
2.
1.
FIGURE 6.12
Disconnect the system from the power source
(outlet or remove battery module if installed) before
attempting any maintenance, installation, removal, or
replacement procedures to prevent electrical shock
and possible damage to system.
Be careful not to damage the contacts of the
sEMG Module, Therapy System, or Module Stim
Board contacts.
When installing sEMG, verify that the model
number is capable of supporting the functions of
sEMG.
•
•
•
6.2
S
EMG MODULE INSTALLATION AND
REMOVAL VECTRA GENISYS ONLY
NOTE:
Twist the Screwdriver to break tabs free.
Carefully remove tabs so as not to cause any
damage to the PC Board Contacts below
the tabs. If installing on the Channel 3/4
Electrotherapy Module, remove tabs from
the Module in the same manner.
Using a Flat Blade Screwdriver, carefully
remove the four breakout tabs on the
System bottom.
See Figure 6.12
.
2.