Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
5 TROUBLESHOOTING
FIGURE 5.33
FIGURE 5.32
FIGURE 5.31
LEAD WIRES TO THEIR
RESPECTIVE COLORS
ON ATTENUATOR
If icons are present, connect known good
sEMG Lead Wire to Channels 1 and 2.
See
Figure 5.31.
NOTE:
Only one Channel at a time can be tested for
sEMG.
Connect the Channel 1 sEMG lead wires into
the Attenuator. Make certain each sEMG Lead
is connected to its respective color on the
Attenuator.
See Figure 5.32.
Press the sEMG Button on Home Screen.
Channel 1 should read 7 or less.
See Figure
5.33.
If Channel 1 reads less than 7, repeat
steps
2 through 5
on Channel 2. If any Channel
being tested reads greater than 7, replace the
sEMG Module and re-test.
NOTE:
The reading on the Channel not being tested
may vary in its reading. This is insignificant as
it is not under load.
2.
3.
4.
5.
5.19 SEMG AND SEMG + ELECTRICAL
STIMULATION
TESTS