68
Vectra Genisys®/Intelect Legend XT®/ Intelect® Vet Therapy Systems
6 REMOVAL/REPLACEMENT
6.4 THERAPY SYSTEM FAN
A. Tools and Equipment Required
#1 Phillips Screwdriver
B. Therapy System Fan Removal
Separate Top from Bottom.
Refer to 6.3,
part B.
Using a #1 Phillips Screwdriver, remove the
Two Fan Retaining Screws securing the Fan
to the System Top.
See Figure 6.22.
Remove the Fan Harness from the Control
Board.
See Figure 6.23.
Remove the Fan Baffle from the Fan Housing.
See Figure 6.24.
C. Replacing Fan
Replace new Fan, part number 27158 by
reversing the
steps 1- 3 above.
NOTE:
Do not over tighten the screws. Over
tightening will damage the threads of the
brass standoffs.
1.
2.
3.
4.
1.
FIGURE 6.22
FIGURE 6.23
REMOVE
SCREWS
REMOVE
FAN HARNESS
FIGURE 6.24
REMOVE
BAFFLE
68
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.