3 Proper Operation
12
3.2 Leveling and Location
1) Location: This scale must be placed on a flat and stable surface. Please keep the scale away
from the direct path of oscillating fans, ventilation systems, or strong drafts as these air disturbances
can be picked-up by the scale’s very sensitive weighing platform and may cause incorrect weight
readings.
2) Leveling: If the scale is not properly leveled, please adjust the 4 adjustable legs at the bottom of
the scale. Turn the legs clockwise or counterclockwise so as to center the bubble of the leveling
gauge inside the indicated circle. Turning the adjustable legs counter-clockwise (viewed from top of
scale) will lower that part of the scale. Turning the adjustable legs clockwise (viewed from top of
scale) will raise that part of the scale. (See Fig.)