25
CAL 7 Weight Measurement Mode using an Equivalent Circuit
Display or Key Used
Description
CAL 1
Enter the maximum capacity allowed
indicated on the label of the load cell
EX) Weight: 200kg
CAL 2
Abbreviation of DIVISION, indicating the
division value
EX) The minimum division value 0.05
☞
CAL 3
Start the process without weight on scale
☞
After 3 beeps, a ‘Good’ message is displayed
if normal
CAL 4
The maximum output value of the loadcell
Input as (Rated Output Voltage / mV/V)
EX ) For 2mV/V, enter as 2.00000
☞
If entered value is normal, Good
PASS
CAL 5
Displayed in a normal progress
☞
After the display from CAL.END
BYE… moves to the Weighing Mode