General Function
F01
Function
Set-up for the Weighing Unit (Weighing Unit)
Display Signification
F01. 0
Kilogram (kg)
F01. 1
Gram (g)
Set-up Range
(0, 1)
F01. 2
Ton (t)
F02
Function
Set-up for the Load Cell Type (Tension or Compress, Tension & Compression)
Display Signification
F02. 0
Load cell in compression type or tension type
Set-up Range
(0, 1)
F02. 1
Load cell in both compression and tension type
* When A_UP S/W3 is ON, the value of Zero increases
F03
Function
Set-up for the weight indication rate ( FND display indication rate)
Display Signification Display Signification
F03. 0
200 times / sec
F03. 5
16 times / sec
F03. 1
100 times / sec
F03. 6
18 times / sec
F03. 2
80 times / sec
F03. 7
4 times / sec
F03. 3
64 times / sec
F03. 8
2 times / sec
Set-up Range
(0 ~ 9)
F03. 4
32 times / sec
F03. 9
1 time / sec
F04
Function
Set-up for the Digital Filter
Display Signification
Set-up Range
(1 ~ 49)
F04. 1 ~ 49
1: Places with low levels of vibration
~ 49: Places with high levels of vibration
25 in case of general weighing operation
F05
Function
Set-up for the Stability Condition of weight
Display Signification
F05. 0
Stable lamp is on if there is a variation in weight within 0.5
divisions in the designated time.
F05. 5
Stable lamp is on if there is a variation in weight within 2.5
divisions in the designated time.
Set-up Range
(0 ~ 9)
F05. 9
Stable lamp is on if there is a variation in weight within 5
divisions in the designated time.
Note 1. This function recognizes it as stable if the variation in weight does not exceed the set range of A/D
Count within the designated time. If the Indicator is located in a place with high levels of vibration,
then increase the number, and if the Indicator is located in a place with low levels of vibration,
then decrease the number in order to quickly reach stability in weight measurement.
Summary of Contents for CI-502A
Page 1: ... 2 ...
Page 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Page 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Page 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Page 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Page 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Page 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Page 83: ... 84 MEMO ...
Page 84: ... 85 MEMO ...
Page 85: ... 86 MEMO ...
Page 86: ... 87 MEMO ...
Page 87: ... 88 ...
Page 88: ... 89 ...