35
Set Mode
(1) How to Enter the Set Mode
Turning on the power while pressing
key on the front of the Indicator starts it in
Set Mode.
To enter into the Set Mode while using the Indicator normally,
Press
key fot approximately 3 seconds.
(2) Key Used
~
: use to change the set value.
: saves the changed set values, and moves to the initial display of the Set Mode.
: cancels and clears the set values, and moves to the initial display of the Set Mode.
(3) Set-up Menu for the Set Value Changes (F01~F99)
*Note. The number in bracket ( ) is the default value.
General Function
F01
(0)
Set-up for the weight unit (0~2)
kg , g , ton
F02 (0)
Set-up for the load cell type(0~1)
(Tension or Compression, Tension &
Compression)
Adjust with A_UP S/W3 ON in case of
General / Tension & Compression
F03
(0)
Set-up for the weight indication rate (0~9)
(FND display indication rate)
200 times / sec ~ 1 time / sec
F04
(25) Set-up for the Digital Filter (01 ~ 49)
Places with low levels of vibration ~ high
levels of vibration
F05
(5)
Set-up for the Stability Condition of
weight(0~9)
Places with low levels of vibration ~ high
levels of vibration
F06 (5)
Set-up for the Automatic Zero Tracking
Compensation (0~9)
Select from ‘do not use automatic
settings’ to differential settings based on
the weight calibration value.
F07
(0)
Weight Backup function(0~2)
0: Do not use Weight Backup
1: Use Weight Backup (calibration)
2: Use Weight Backup (reference:
operation)
F08
(0)
Set the Hold Type (0~2)
Average Hold/ Peak Hold / Sampling
Hold
F09
(30) Set-up for the average hold time (01~99)
0.1 sec ~ 9.9 sec
F10 (1)
Set-up for the operation condition of Zero
key and the Tare key (0 ~ 1)
0: Zero and Tare key always available
1: Zero and Tare key only available when
the weight is stable
Summary of Contents for CI-502A
Page 1: ... 2 ...
Page 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Page 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Page 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Page 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Page 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Page 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Page 83: ... 84 MEMO ...
Page 84: ... 85 MEMO ...
Page 85: ... 86 MEMO ...
Page 86: ... 87 MEMO ...
Page 87: ... 88 ...
Page 88: ... 89 ...