Appendix 3. Error Code Table
(1)Error that can occur in the Weighing Mode
ERROR 01
▣
Reason for Error
Data stored in the internal memory was deleted due to electric shock.
☞
Troubleshoot
Do the necessary designations in the Set Mode.
ERROR Lc
▣
Reason for Error
The load cell is not connected properly or there is a problem in
the A/D Converter.
☞
Troubleshoot
Check the connection between the scale and the unit.
UNPASS
▣
Reason for Error
The initial zero range has exceed the F13 intial zero range in the Set Mode.
☞
Troubleshoot
Check the weight on scale, and change the F13 value range in
the Set Mode.
ERROR 04
▣
Reason for Error
A key was pressed for a long time or there is a problem in the
key part.
☞
Troubleshoot
Do the key test in Test Mode 1, and if this message continues to appear without any
actual problems, contact the A/S department of our company.
OVER
▣
Reason for Error
A key was pressed for a long time or there is a problem in the
key part.
☞
Troubleshoot
Do the key test in Test Mode 1, and if this message continues to appear without any
actual problems, contact the A/S department of our company.
Summary of Contents for CI-502A
Page 1: ... 2 ...
Page 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Page 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Page 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Page 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Page 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Page 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Page 83: ... 84 MEMO ...
Page 84: ... 85 MEMO ...
Page 85: ... 86 MEMO ...
Page 86: ... 87 MEMO ...
Page 87: ... 88 ...
Page 88: ... 89 ...