`
19
* The Relationship Between the Load Cell Output and Input Sensitivity
The input sensitivity of this product is over 0.2uV/digit maximum.
The system design must satisfy the following formula:
Example #1)
The number of LoadCells: 4 EA
LoadCell Max.Capacity: 500 Kg
LoadCell Rate Output: 2mV/V
1 Division: 0.10 Kg
Load Cell excitation Voltage: 10V (= 10,000 mV)
According to the equation
( 10000 mV* 2mV * 0.1Kg)/(500Kg * 4) = 1
≥
0.2uV
Since the calculated value is higher than 0.2uV, this weight system design does not have any
problems.
Note 5: Internal LOAD CELL Output Switch (SW2)
Turning on the A_UP S/W2 in the interior of the indicator will increase the 1mV/V Zero
Value.
●
In the Test Mode 3 LoadCell Test Mode, mV/V value can be checked by pressing the
Clear key.
●
Set the F02 to 1, and turn on the A_UP S/W2 to use compression and tension
simultaneously.
LoadCell excitation Voltage x LoadCell Output Voltage
x 1 Division Value
0.2uV
≤
------------------------------------------------------------------
LoadCell Max.Capacity x The number of Load Cells
Summary of Contents for CI-502A
Page 1: ... 2 ...
Page 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Page 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Page 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Page 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Page 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Page 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Page 83: ... 84 MEMO ...
Page 84: ... 85 MEMO ...
Page 85: ... 86 MEMO ...
Page 86: ... 87 MEMO ...
Page 87: ... 88 ...
Page 88: ... 89 ...