20
Calibration Mode
What is Calibration?
It refers to adjusting the indicated numerical value of the weight to represent the
actual weight.
(1) How to enter Calibration Mode
Loosen the bolts at the Rea
①
r panel of the Indicator and set the CAL
switch to the ON position as shown in the figure above.
CAL is at the ON position originally.
※
Close the Rear panel of the Indicator, and while pressing the
②
(3 Tare) key,
turn on the power.
Calibrate.
③
④
Upon completion of calibration, set the CAL switch to the OFF position, assemble,
and seal.
If it is calibrated with the CAL switch in the OFF position, SLOC will be displayed.
⑤
Refer to the
※
sealing method on pg. 75