Chapter 5 User Maintenance
Revision Q
HF Series X-ray Generators - Operator’s Manual
5-8
Quantum Medical Imaging, LLC
•
Back Up Time/MAS
setting: Displays current AEC exposure back up
mAs setting (only functions on systems equipped with QG-AEC option).
The system allows for selection of a mAs setting between 10 and 600
mAs (in 10 mAs increments).
•
Generator Timeout
: Sets the delay time for generator to automatically
shut down if no keys are pressed on operator control panel (selectable in
30 minute increments; between 0 minutes (automatic shutdown
disabled) up to 180 minutes).
•
Date Format
setting: Sets date/time format (options are "US" or
"International)