Chapter 4 APR Editor
Revision Q
HF Series X-ray Generators - Operator’s Manual
4-8
Quantum Medical Imaging, LLC
Removing an Existing Character
a. Use the up and down keys next to the
Sentence Position
field to
select the character to remove from the text string. For example, if
the text string "SKULL" is displayed in the center display box, choos-
ing character position number "0" selects the letter "S" for removal,
position "1" selects the letter "K" for removal, and so on. Note that
the character in the selected position appears in the
Character
field.
b. Pressing the key next to the
Delete Character
field deletes the
selected character. Note that when the character is deleted, any
characters to the right of the deleted character will shift to the left
one space thereby "closing up" the deleted character’s space.
Replacing an Existing Character
a. Use the up and down keys next to the
Sentence Position
field to
select the character to replace. For example, if the text string
"SKULL" is displayed in the center display box, choosing character
position number "0" selects the letter "S" for replacement, position
"1" selects the letter "K" for replacement, and so on. Note that the
character in the selected position appears in the
Character
field.
b. Use the up and down keys next to the
Character
field to scroll
through all available alphanumeric characters (beginning with a "no
character" or blank space) until the desired character appears in the
Character
field window.
c. To replace the character in the position chosen in the previous step,
press the key next to the
Replace Character
field. The new char-
acter will now appear in the selected position of the text string.
3. When done editing the Region or View name, you may select
Previous,
Save
to save the new settings and return to the previous screen (Region Edit
or View Edit screen) or select
Previous, No Save
to ignore the changes
and return to the previous screen (Region Edit or View Edit screen). To
restore region (or view) name to the factory default settings, press the key
next to the
Factory Reset
field.
Editing a View Name
Subordinate to each region, there may exist up to 10 individual "view
names". To change the name of a view or create a new view (where
there is currently a blank field), proceed as follows:
a. Access the APR Region Edit screen as described previously
(see Figure 4-6).
b. In the APR Region Edit screen, press the key adjacent to the
Change View Settings
field. The APR View Selection
screen appears:
Summary of Contents for Quantum Odyssey HF QG-32
Page 7: ...Chapter Chapter 1 1 1 1 SAFETY NOTICES SAFETY NOTICES 1 1...
Page 8: ...1 2...
Page 21: ...Chapter 2 1 GENERAL INFORMATION 2...
Page 22: ...2 2...
Page 39: ...Chapter OPERATION 3 1 3...
Page 40: ...3 2...
Page 67: ...Chapter APR EDITOR 4 1 4...
Page 68: ...4 2...
Page 83: ...Chapter Chapter 5 1 5 1 USER USER MAINTENANCE MAINTENANCE 5 5...
Page 84: ...5 2 5 2...
Page 97: ...Chapter 6 1 WARRANTY INFORMATION 6...
Page 98: ...6 2...