Chapter 3 Operation
HF Series X-ray Generators - Operator’s Manual
Revision Q
Quantum Medical Imaging, LLC
3-5
DAILY TUBE WARM-UP PROCEDURE
All tube manufacturers recommend X-ray tube seasoning following installation
and daily prior to use. Systems used infrequently should have the tube seasoned
on a daily basis. It is important that these procedures be performed to maintain
both the tube manufacturer's and Quantum Medical Imaging, LLC’s warranties. A
typical daily seasoning procedure is provided below. If the system has not been
used for several days, or upon installation, refer to the tube manufacturer's
instructions. Proceed as follows:
1. Select MANUAL mode from the Operator Control Panel.
2. Take exposures listed in Table 3-1 using a 200L mA focal spot and allow 30-
seconds between exposures.
3. The system is ready for operation.
NOTE: "Exposure 8" X-ray Tube seasoning set-
tings applies only to generators capable of 150kV.
Table 3-1: Daily X-ray Tube Seasonin g Exp osu res
EXPOSURE
KVP
TIME
(SECONDS)
1
50
0.1
2
60
0.1
3
70
0.1
4
80
0.1
5
90
0.1
6
100
0.1
7
110
0.1
8
120
0.1
Summary of Contents for Quantum Odyssey HF QG-32
Page 7: ...Chapter Chapter 1 1 1 1 SAFETY NOTICES SAFETY NOTICES 1 1...
Page 8: ...1 2...
Page 21: ...Chapter 2 1 GENERAL INFORMATION 2...
Page 22: ...2 2...
Page 39: ...Chapter OPERATION 3 1 3...
Page 40: ...3 2...
Page 67: ...Chapter APR EDITOR 4 1 4...
Page 68: ...4 2...
Page 83: ...Chapter Chapter 5 1 5 1 USER USER MAINTENANCE MAINTENANCE 5 5...
Page 84: ...5 2 5 2...
Page 97: ...Chapter 6 1 WARRANTY INFORMATION 6...
Page 98: ...6 2...