Chapter 4 APR Editor
Revision Q
HF Series X-ray Generators - Operator’s Manual
4-4
Quantum Medical Imaging, LLC
Figure 4-1 provides an overview for accessing the various APR edit mode screenss.
Each screen is described in detail in the paragraphs that follow.
ACCESSING THE APR EDIT MODE
APR Edit mode is accessed through the Service Mode. To gain access to the Ser-
vice Mode, password entry is required. This is to prevent unintended and/or
unauthorized changes to system and/or APR screen settings. To access the APR
Edit mode, proceed as follows:
1. Set the ON/OFF circuit breaker CB1 on the generator cabinet to ON position.
2. Set the power on/standby switch on the Operatorback of the E-ZVET Control
Panel (OCP)to the standby (
) position, then to the on ( ) position.
3. The system begins the start up sequence (the "Welcome Screen" is displayed
and "Testing. Standby..." is flashing) as shown in Figure 4-2.
4.
Welcome Screen
Within five seconds of start up, press the upper left field
select key as shown in Figure 4-3. The Password menu displays:
Figure 4-2. Passwor d Menu
PRESS TO
ACCESS
PASSWORD
MENU
FIELD SELECT KEYS