CR System Service Procedures
7–10
6K7304 | 2014-10-20
Install the SD Card
1. Insert the new SD card into its slot in cable compartment. Pay attention to the SD card
direction:
2. Connect the power supply cable and USB cable to the scanner
3. Install the cable compartment cover. See
Install the Cable Compartment Cover
4. Connect the USB cable to the work station computer.
5. Connect the power supply cable to the Mains wall socket.
6. Press the On/Off button to switch on the scanner.
Confirm the Scanner Function after Replacing the SD Card
1. Perform the System Self Test to make sure system functions properly. See
Run System
Self Test
2. Perform the Acceptance Test Procedure to make sure the image acquisition functions
properly. See
Acceptance Test Procedure
3. Open your imaging software.
4. Scan an unexposed 35.6 x 43.2 cm (14 x 17 in.) cassette, or if unavailable, any other
size cassette.
5. Verify that the cycle completes without error.