CR System Service Procedures
7–4
6K7304 | 2014-10-20
4. Holding the plastic handles, remove the Screen Transportation Unit.
Caution
Do Not insert your fingers between the cassette opening and the door.
Do Not get your fingers caught between the loader and the screen transportation unit
Install the Screen Transportation Unit
1. Insert the Screen Transportation Unit, and secure the two tightening knobs.
2. Connect the power supply cable to the Mains wall socket.
3. Switch on the system and verify that the LED on the Scanner is green and that your
imaging software shows that the Scanner status is
OK
.
Confirm the Scanner Function after Replacing Screen Transportation
Unit
1. Perform the System Self Test to make sure system functions properly. See
Run System
Self Test
2. Perform the Acceptance Test Procedure to make sure the image acquisition functions
properly. See
Acceptance Test Procedure
3. Open your imaging software.
4. Scan an unexposed 35.6 x 43.2 cm (14 x 17 in.) cassette, or if unavailable, any other
size cassette.
5. Verify that the cycle completes without error.
Summary of Contents for Carestream Vita Flex CR System
Page 1: ...l Publication No 6K7304 2014 10 20 Version 2 0 CARESTREAM Vita Flex CR System User Guide...
Page 6: ...iv 6K7304 2014 10 20...
Page 8: ...About this Guide 1 2 6K7304 2014 10 20...
Page 14: ...Site Preparation 2 6 6K7304 2014 10 20...
Page 32: ...Install the CR System 4 16 6K7304 2014 10 20...
Page 40: ...Use the CR System 5 8 6K7304 2014 10 20...
Page 46: ...Use the CR System 5 14 6K7304 2014 10 20...
Page 56: ...Maintain and Clean the System Cassette Screen Roller Cleaning 6 10 6K7304 2014 10 20...
Page 82: ...System Self Test 9 6 6K7304 2014 10 20...