Maintain and Clean the System (Cassette, Screen & Roller Cleaning)
6–4
6K7304 | 2014-10-20
5. Click
OK
to approve the performed procedure and to reset the counters and then exit
the
Utilities Center
software.
Note
If you click
Cancel
, the counters will not be reset and cleaning messages will appear for the next
100 cycles.
Confirm the Rollers Function
1. Switch on the system and verify that the LED on the Scanner is green and that your
imaging software shows that the Scanner status is
OK
.
2. Insert an unexposed
35.6 x 43.2 cm (14 x 17 in.)
cassette, or if unavailable, any other size
cassette and perform a scan.
3. Verify that the cycle completes without error.