Install the CR System
6K7304 | 2014-10-20
4–13
Acceptance Test Procedure
The Acceptance Test Procedure (from this point on will be referred as ATP) confirms that
the CR system was not damaged during shipment and is ready to begin patient testing.
Note
This test is mandatory. If not performed, a warning message will appear requesting to perform
the test.
Access ATP
1. From the desktop of the Image Suite computer, click the I
mage Suite 4.0
icon to
open the Image Suite application.
2. In the Login window that appears, type
User name
(
admin
) and
Password
(
123456
).
3. On the header bar of the Image Suite application, click the
Help
menu button.
4. From the displayed Help menu, select
Launch Acceptance Test
.
Summary of Contents for Carestream Vita Flex CR System