Maintain and Clean the System (Cassette, Screen & Roller Cleaning)
6–6
6K7304 | 2014-10-20
3. Push the slider on the cassette upwards.
A centimeter of the screen sticks out of the cassette.
4. Gently pull out the rest of the screen.
Clean the GP-2 Phosphor Screens
Under normal use, the GP-2 Phosphor Screens will eventually show wear. Screen wear can
result in image artifacts. This wear may occur from abrasion of the protective overcoat or
damage to the surface of the screen.
Prerequisites:
Read and follow the instructions in the manufacturer’s Material Safety Data Sheet
(MSDS) prior to use.
Caution
Cleaning materials other than those recommended can contain chemicals that cause visible or
hidden damage to the screen or cassette and could result in immediate or future image artifacts.
Never use isopropyl alcohol (isopropanol, rubbing alcohol) to clean screens or cassettes.
Cleaning Materials
•
Several soft, lint-free cloths
•
Carestream X-OMAT Screen Cleaner CAT number 845-4977 (473 mL/ 16 Fluid
Ounces) or CAT number 103-0428 (177 mL/ 6 Fluid Ounces).
Summary of Contents for Carestream Vita Flex CR System
Page 1: ...l Publication No 6K7304 2014 10 20 Version 2 0 CARESTREAM Vita Flex CR System User Guide...
Page 6: ...iv 6K7304 2014 10 20...
Page 8: ...About this Guide 1 2 6K7304 2014 10 20...
Page 14: ...Site Preparation 2 6 6K7304 2014 10 20...
Page 32: ...Install the CR System 4 16 6K7304 2014 10 20...
Page 40: ...Use the CR System 5 8 6K7304 2014 10 20...
Page 46: ...Use the CR System 5 14 6K7304 2014 10 20...
Page 56: ...Maintain and Clean the System Cassette Screen Roller Cleaning 6 10 6K7304 2014 10 20...
Page 82: ...System Self Test 9 6 6K7304 2014 10 20...