Install the CR System
4–4
6K7304 | 2014-10-20
Hardware Installation
Prerequisites:
Make sure that the power source for the system is in the range of 100V AC- 240V AC 1.5 A
Note
For optimal results it is recommended that the light level in the room is less than 500 lux to
minimize the imaging plates exposure to light. Avoid placing the scanner beneath windows or
sources of direct, strong light
For more information on
ambient operating condition requirements, see the VITA FLEX
Safety and Regulatory Guide
Important
The CR system should be positioned so that there is always easy access to the mains power supply
1. Place the CR System on a stable flat surface and turn it onto its back to gain access to
the service door at the rear of the scanner.
2. Open and remove the cable compartment cover located in the rear of the scanner.
Use the cover removal tool.
3. Connect the main power cable and the USB cable to the scanner and close the cable
compartment cover.