CR System Service Procedures
6K7304 | 2014-10-20
7–5
Replace the Power Supply and USB cables
Remove the Power Supply Cable
1. Press the On/Off button to switch off the scanner and ensure that the LED power
indicators are off.
Note
You must make sure that the LED power indicators are off before any service or maintenance
procedures.
2. Unplug the power supply from the Mains wall socket.
3. Turn the scanner onto its back.
4. Remove the cable compartment cover at the rear of the CR System. See
Remove the
Cable Compartment Cover
Caution
Check that the power supply is not hot before touching it. If the power supply malfunctions, it
may get very hot.
5. Disconnect the power supply cable and the USB cable.