Appendix D Fault Management
59
675–101–101 Rev. P
E## Fault condition
Consequence
Clas-
sification
Software
Response
Corrective
Action Required
E10 Non-volatile
memory fault
Unable to
retrieve/set unit
configuration and
calibration data
Unusable
User lockout:
Error "E##"
prompt
Service: Fix or
replace PCB
E11 Calibration data
lost
Sensor readings
invalid
Unusable
User lockout:
Error "E##"
prompt
Service: Low
level calibration
(O
2
, Pressure
and Flow)
E12
Configuration DIP
settings and/or
PT PRESENT
different to non-
volatile
configuration
record
Possible
incomplete unit
set-up performed
Unusable
User lockout:
Error "E##"
prompt
Service: Perform
set-up procedure
E20
Charged battery
voltage too low
(<11V) when
under test load
Battery capacity
low
No backup
Battery fault icon
flashes and "E##"
alarm
Service: Fix
Battery or
charger
E21
External supply
voltage too low
(<14V) to charge
battery (battery
flat)
Battery will not
charge
No backup
Battery low alarm
continues even
through external
power applied -
"E##" alarm
User: Plug in
correct external
supply
Service: Fix PSU
circuits
E22 Analogue supply
rails out of limits
Unreliable sensor
readings
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
circuits
E23
Valve driver
supply rails out of
limits
Valve operations
unreliable
Unusable
User lockout:
Error "E##"
prompt
Service: Fix valve
supply rail
E24
Hardware 'safe-
start' watchdog
disabled
Valves disabled
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
reset/safe-start
circuits
E30
Pressure sensor
fault (ADC hits
rail)
Pressure sensor
readings invalid
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
sensor/circuits
E31
Zero valve not
connected (via
sense)
Pressure sensor
readings
unreliable
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
valve/circuits
E32
Zero valve
activation fault
(via sense)
Pressure sensor
readings
unreliable
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
valve/circuits
E33
Unable to auto-
zero pressure
sensor
Pressure sensor
readings
unreliable
Unusable
User lockout:
Error "E##"
prompt
Service: Fix
valve/sensor/circ
uits
E41
Dump valve not
connected (via
sense)
No over pressure
protection
Restricted Restricted mode:
Error "E##" alarm
Service: Fix
valve/circuits
E42
Dump valve
activation fault
(via sense)
No over pressure
protection
Restricted Restricted mode:
Error "E##" alarm
Service: Fix
valve/circuits
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...