54
Appendix C Alarm Troubleshooting
675–101–101 Rev. P
Alarm
Priority
Possible Cause
Actions
Battery fault
High (Cannot be
reset)
•
Battery
disconnected
•
Battery failing to
hold charge
•
Push Alarm Mute button for 3
seconds to silence alarm
•
Refer to Service Technician.
Low battery
voltage
< 11.10 V for
5 seconds).
Medium
•
Battery
disconnected
•
Battery failing to
hold charge
•
Push Alarm Mute button for 3
seconds to silence alarm
•
Connect AC power
AC power
disconnected
Low
•
AC power
disconnected
•
Push Alarm Mute button for 3
seconds to silence alarm
•
Reconnect the AC power.
High
NCPAP /
Pres Low
(CPAP > 3
cmH
2
O
above set
for 15
seconds).
High
•
NCPAP / Pres Low
setting change
•
Circuit disconnect /
reconnect
•
Push Alarm Mute button for 3
seconds to silence and reset alarm
limits
•
Reset alarm limits after setting
change and patient circuit
disconnect / reconnect
Low NCPAP
/ Pres Low
( CPAP < 2
cmH
2
O
below set
for 15
seconds)
or
< 1.5 cmH
2
O
at any time).
High
•
NCPAP / Pres Low
setting change
•
Circuit disconnect /
reconnect
•
Circuit leak
•
Push Alarm Mute button for 3
seconds to silence and reset alarm
limits
•
Reset alarm limits after setting
change and patient circuit
disconnect / reconnect
•
Check for leaks in patient circuit
High
BiPhasic or
BiPhasic tr*
pressure
(MAP > 3
cmH
2
O
above
set for 15
seconds).
High
•
Pres High setting
change
•
Circuit disconnect /
reconnect
•
Push Alarm Mute button for 3
seconds to silence and reset alarm
limits
•
Reset alarm limits after setting
change and patient circuit
disconnect / reconnect
BiPhasic or
BiPhasic tr*)
mode fails to
operate as
set.
High (Cannot be
silenced)
•
See description of
error code
displayed
•
Revert to nCPAP mode
•
Refer to Service Technician
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...