Chapter 4 Unpacking and Setup
17
675–101–101 Rev. P
User Verification Test
WARNING
Do not attach Generator to the patient until User Verification and initial
setup into NCPAP mode is complete.
CAUTION
Although failure of any of the user verification tests described in this section will not
prevent the ventilator from functioning, the ventilator should be checked to make
sure it is operating correctly before use on a patient.
Power-on Check
This test is run automatically on power up of the driver and automatically performs
the following checks:
•
Flash ROM
•
Hardware Inputs/Outputs
•
Audible and visual alarms indicators
•
Test and calibration of pressure sensor
•
Test of dump valve
The unit carries out a full functional check during this time. If unsuccessful, the
screen remains darkened and the warning bar remains on. In this case, check for the
following;
•
Power Supply not connected
•
Battery voltage low
If the checks are successful, the screen changes to Power Up Screen.
During the Power Up check:
•
Screen image shown in negative
•
Warning bar comes on for one second
•
Transducer Assembly LED comes on for one second
•
Audible alarm sounds for one second
•
Dump valve is tested
•
Pressure is set to zero
After two seconds screen changes to Setup Screen. Alarm limits are disabled and a
flashing question mark appears under the NCPAP / Pres Low flow meter screen
indicator.
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...