Chapter 5 Operation
29
675–101–101 Rev. P
2. Alarm set/confirm Screen
Figure 10 Alarm set/confirm Screen
Touch the NCPAP button or the Alarm Mute / Reset button for 3 seconds to set the
alarms and to move to the next screen. If either button is not touched within 2
minutes, the alarm limits will be set automatically. When the alarm limits have been
set, the screen display changes to the Mode Select Screen with the driver operating
in NCPAP mode.
Note:
Press High Flow meter must be checked through the Start-up screen and be set
during operation. This enables manual breath or back-up apnea breath, where
applicable, to be active.
CAUTION
Do not adjust any of the Alarm Limit settings to an extreme value. Selecting an
extreme value can make the alarm system inadequate for the patient.
Figure 11 Mode Select Screen
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...