Chapter 5 Operation
33
675–101–101 Rev. P
Every hour
Repeat the checks listed above in Final Check.
Every 3-4 hours
Loosen the generator straps and release the tubes from the central Velcro strip. The
nasal area can be cleaned with warm sterile water. Do not apply creams or
ointments.
Ensure that:
•
Nasal prongs/mask is not occluded with mucus/water droplets
•
Patient prongs/mask and bonnet continue to fit appropriately.
•
Re-apply the generator as described above.
Power Off
To power off:
1. Disconnect infant from SIPAP driver.
2. Turn off SIPAP driver.
If therapy has been discontinued, after turning off power:
1. Turn the flow rate of the low pressure flow meter and the high pressure flow to
zero.
2. Disconnect air and oxygen high pressure hoses from wall outlet.
3. Remove nCPAP circuit and generator and dispose of per hospital procedure.
Directions for using the AirLife Infant nCPAP
System
Please refer to P/N 36-5569 included in the AirLife Infant nCPAP Fixation Device for
the
Directions for Use
. (The AirLife Infant nCPAP system is only available in the
United States and Canada.)
WARNING!
CareFusion consumables are specifically designed to be used with Infant
Flow Drivers and are the only consumables validated for use with Infant
Flow devices.
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...