Chapter 5 Operation
27
675–101–101 Rev. P
Parameter Default Value on Change of Mode
Some controls are active in more than one operating mode. In these instances, there
is a separate default value for operating modes as illustrated with the following table.
Settings that are changed by the operator in one specific mode will be maintained if
the mode is changed to another mode within the same mode group. All defaults shall
revert to factory default on power-cycling or software restart.
Setting a Manual Breath
The manual-breath function is available in CPAP, Biphasic, and Biphasic tr modes.
For manual breath to be active when the manual button is selected, the pre-use
pressure-high check has been completed and the pressure high-flow meter is set for
preferred manual breath. One manual breath is delivered per button-press.
Note:
The Pressure High Flow Meter must have flow above zero in order to deliver manual
breaths.
Table 2 Parameter Default Value
Mode
Parameter
NCPAP
NCPAP + rate
monitoring
BiPhasic
Bi rate
monitoring
BiPhasic tr
T-High
Default for NCPAP apply
Default for BiPhasic apply
Default for
BiPhasic tr
Rate
Default for BiPhasic apply
Rb
Default for
BiPhasic tr
T
apnea
/ T
LBR
This setting applies to all modes: system-wide default applies to all modes
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...