Infant Flow SiPAP
1
675–101–101 Rev. P
Chapter 1 Product Description
Infant Flow
™
SiPAP driver provides a non-invasive form of respiratory support
designed for infants in hospital environments such as Neonatal and Pediatric
Intensive Care Units. It can also be used when transporting these patients within the
hospital environment.
Infant Flow
SiPAP driver is currently available in a Plus or Comprehensive*
configuration. The Plus configuration provides NCPAP and time triggered BiPhasic
modes with and without breath rate monitoring. The Comprehensive* configuration
offers these features plus a patient triggered BiPhasic mode with apnea backup
breaths. The Infant Flow
SiPAP driver comes standard in all configurations with an
LCD touch screen display, pressure time waveform graphics, integrated patient
monitoring, alarms for high and low pressure and FiO
2
and up to 2 hours of backup
battery power.
As a result of the unique, patented design, the Infant Flow, Infant Flow LP or AirLife
Infant nCPAP System Generators have been proven to provide the most stable
CPAP at the lowest work of breathing for patients compared to other devices
(1)
. The
outstanding performance of variable flow generators is irrespective of patient
demand or expiratory flows. The SiPAP driver has been designed and tested to
perform optimally when used only with accessories available from CareFusion.
These accessories include the variable flow generator, single limb breathing circuit,
nasal prongs, nasal masks, and fixation device, either bonnet or headgear.
Infant Flow SiPAP Features
The expanded capabilities of the Infant Flow
SiPAP Plus and Comprehensive*
configurations allow for applications to broader range of patients who may otherwise
not be candidates for non-invasive respiratory support from NCPAP alone
(2,3)
.
NCPAP –
continuous positive airway pressure based on clinician set pressure.
Breath rate monitoring/alarm can be activated in this mode.
BiPhasic
– time triggered pressure assists are delivered based on clinician set
Time-High, rate and pressure criteria. Breath rate monitoring/alarm can be activated
in this mode.
BiPhasic tr*
– patient triggered pressure assists delivered based on clinician set
Time-High and pressure criteria. Breath rate monitoring/alarm and Apnea backup
breaths are automatically active in this mode.
Infant Flow, Infant Flow LP, or AirLife Infant nCPAP System Generator
– The
Infant Flow
Generator is a fluidic device for the generation of consistent infant nasal
CPAP with a low work of breathing compared to other devices
(1)
.
Fully integrated alarm package
– Supply gases failure, High Patient Pressure, Low
patient pressure, high and low delivered Oxygen concentration, change from AC to
DC power source, low and flat battery charge status and Low breath rate/apnea
alarm.
Summary of Contents for Infant Flow SiPAP
Page 1: ...Infant Flow SiPAP Operator s Manual ...
Page 10: ...x Infant Flow SiPAP 675 101 101 Rev P Electromagnetic Environment Specifications ...
Page 11: ...Infant Flow SiPAP xi 675 101 101 Rev P ...
Page 14: ...xiv Infant Flow SiPAP 675 101 101 Rev P Infant Flow SiPAP ...
Page 20: ...6 Chapter 2 Product Specifications 675 101 101 Rev P Infant Flow SiPAP ...
Page 38: ...24 Chapter 5 Operation 675 101 101 Rev P Rear Panel Figure 8 Rear Panel ...
Page 48: ...34 Chapter 5 Operation 675 101 101 Rev P Infant Flow SiPAP ...
Page 64: ...50 Appendix A Product Configurations 675 101 101 Rev P Infant Flow SiPAP ...
Page 65: ...Infant Flow SiPAP 51 675 101 101 Rev P Appendix B Pneumatic Diagram ...
Page 66: ...52 Appendix B Pneumatic Diagram 675 101 101 Rev P Infant Flow SiPAP ...
Page 70: ...56 Appendix C Alarm Troubleshooting 675 101 101 Rev P Infant Flow SiPAP ...
Page 76: ...62 Appendix D Fault Management 675 101 101 Rev P Infant Flow SiPAP ...