Chapter 4 Unpacking and Setup
675–101–101 Rev. P
Figure 6: Flow pressure nomogram for 675-CFG-XXX when used with the Infant Flow LP or
AirLife variable-flow generator
Individual devices have a tolerance of up to ±15 percent from that illustrated in the
nomogram and, in particular, at pressures below 2cmH
2
O.
Flow Pressure Nomogram (for
reference only) to show typical flow
pressure relationships. This is not
meant to establish actual product
performance.