Section 8. Operation
405
where
percent-of-
reading = 1050 mV • ±0.04%
= ±0.42 mV
and
offset = 10 µV.
Therefore,
accuracy = ±0.42 mV + 10 µV
= ±0.52 mV
Electronic Noise
Electronic "noise" can cause significant error in a voltage measurement,
especially when measuring voltages less than 200 mV. So long as input
limitations are observed, the PGIA ignores voltages, including noise, that are
common to each side of a differential-input pair. This is the common-mode
voltage. Ignoring (rejecting or canceling) the common-mode voltage is an
essential feature of the differential input configuration that improves voltage
measurements.
Figure
PGIA with Input Signal Decomposition
(p. 418),
illustrates the
common-mode component (V
cm
) and the differential-mode component (V
dm
) of a
voltage signal. V
cm
is the average of the voltages on the V+ and V– inputs. So,
V
cm
= (V+ + V
–)/2 or the voltage remaining on the inputs when V
dm
= 0. The
total voltage on the V+ and V–
inputs is given as V+ = V
cm
+ V
dm
/2, and V
L
=
V
cm
– V
dm
/2, respectively.
8.1.2.2 Thermocouple Measurements — Details
Thermocouple measurements are special case voltage measurements.
Note
Thermocouples are inexpensive and easy to use. However, despite
the use of a thermocouple in
Quickstart
(p. 39),
the CR6 is not designed for
accurate measurement when thermocouples are attached directly to the
wiring panel.
CR6 design features that cause thermocouple measurement inaccuracy include:
•
lack of an insulating wiring-terminal cover.
•
no high-thermal mass element incorporated in the wiring panel.
•
position of the on-board reference thermistor in the wiring panel is not
optimal.
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......