Section 7. Installation
304
Dwell Burst Measurement
'This program makes 34802 measurements of two single-ended channels at
'93750 Hz. Sample pattern is 1,1,1..., pause, 2,2,2..., pause.
'Measurement cycle is repeated every 2 Sec. The following programming features are
'key to making this application work:
'--PipelineMode.enabled.
'--Dash (-) placed before channel number.
'--Measurement count per channel set with VoltSE() Count=
34802
.
'--Measurement speed set with VoltSE()
[SampleRate (Hz)]=93750
.
'--Scan() BufferOption increased to 5.
'NOTES:
'--Sampling occurs at the beginning of the Scan() interval.
'--All measurements for one channel are placed in a single large variable array.
'--The large array is stored in a single long record in the data table.
'--The exact sampling interval is calculated as follows:
' SampleTime = INT(MuxTime *( Reps + 1) + 2 ms). MuxTime=multiplexed measurement time.
PipeLineMode
Public
DwellBurstSE1(34802)
Public
DwellBurstSE2(34802)
DataTable
(DwellBurstSEData,1,-1)
Sample
(34802,DwellBurstSE1(),FP2)
Sample
(34802,DwellBurstSE2(),FP2)
EndTable
BeginProg
'Scan(Interval,Units,BufferOption,Count)
Scan
(1,Sec,5,0)
'VoltSE(Dest,Count,Range,-SEChan,MeasOff,SettlingTime,[SampleRate (Hz)],Mult,Offset)
VoltSe
(DwellBurstSE1(),34802,mV200,-U1,False,100,93750,1.0,0)
VoltSe
(DwellBurstSE2(),34802,mV200,-U2,False,100,93750,1.0,0)
CallTable
DwellBurstSEData
NextScan
EndProg
200
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......