Bridge101A
Red 485
Dark Blue Pantone 289
Light Blue 58% of Pantone 289
*Bridge101A Range, Resolution and Accuracy
Nominal Range
±30 mV
±150 mV
±1000 mV
Measurement Range
±30 mV
±160 mV
±1200 mV
Resolution
1 µV
5 µV
50 µV
Calibrated Accuracy
±0.01% FSR;
±3 Microvolts
±0.01% FSR;
±16 Microvolts
±0.01% FSR;
±120 Microvolts
Input Range
0 to 2.5V
0 to 2.5 V
0 to 2.5 V
Reference Voltage
2.5 V
2.5 V
2.5 V
Battery Warning
WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT,
CHARGE, FORCE OVER DISCHARGE, DISASSEMBLE, CRUSH, PENETRATE OR INCINERATE.
BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 80 °C (176 °F).
DOC 1242035-00 REV 16 2015.04.14
Bridge101A-30
30mV Bridge Data Logger
Bridge101A-150
150mV Bridge Data Logger
Bridge101A-1000
1000mV Bridge Data Logger
Description
Bridge101A
Range
See Table Below*
Resolution
Accuracy
Memory
1,000,000 readings; software configurable memory wrap
330,000 readings in multiple start/stop
Wrap Around
Yes
Reading Rate
4Hz to 1 every 24 hours
LED Indicator
Red and Green
Units
V, mV, µV, Engineering Units specified through software
Required Interface
IFC200
Baud rate
115,200
*Battery Life
10 months typical at a 1 minute reading rate with a 350
Ω
load
2 years typical at a 1 minute reading rate with a 1000
Ω
load
Operating Environment
-40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 90 %RH non-condensing
Material
ABS plastic
Dimensions
1.4 in x 2.5 in x 0.6 in (36 mm x 64 mm x 16 mm)
Weight
0.8 oz (24 g)
Approvals
CE
Product Information Card
*MadgeTech recommends using 350 or 1000
Ω
strain gages. 120
Ω
strain gauges can be used with the Bridge101A but
it is not recommended as it will reduce the battery life, potentially increase noise level and limit the low end temperature
operation.
www.calcert.com
1.888.610.7664