SECTION 13. CR23X MEASUREMENTS
13-19
maximum and additive. A temperature of 45°C
is measured with a type T (copper-constantan)
thermocouple, using the ±5 mV range. The
nominal accuracy on this range is 2.5 µV
(0.05% of 5 mV), which at 45°C changes the
temperature by 0.06°C. The RTD is 25°C but is
indicating 25.3°C, and the terminal that the
thermocouple is connected to is 0.3°C cooler
than the RTD.
TABLE 13.4-4. Example of Errors in
Thermocouple Temperature
Source
Error °C
% of Total Error
1
o
C
1% Slope
Error Error
Reference Temp.
0.6
36.1
69.6
TC Output
ANSI
1.0
60.1
0.01 x 20°C
0.2
23.2
Voltage
Measurement
0.06
3.6
7.0
Reference
Linearization
0.001
0.1
0.1
Output
Linearization
0.001
0.1
0.1
Total Error
With ANSI error
1.662
100
Assuming 1%
0.862
100
slope error
13.4.2 USE OF EXTERNAL REFERENCE
JUNCTION OR JUNCTION BOX
An external junction box is often used to
facilitate connections and to reduce the
expense of thermocouple wire when the
temperature measurements are to be made at a
distance from the CR23X. In most situations, it
is preferable to make the box the reference
junction, in which case its temperature is
measured and used as the reference for the
thermocouples; copper wires are run from the
box to the CR23X (Section 7.4). Alternatively,
the junction box can be used to couple
extension grade thermocouple wire to the
thermocouples being used for measurement,
and the CR23X panel used as the reference
junction. Extension grade thermocouple wire
has a smaller temperature range than standard
thermocouple wire, but meets the same limits of
error within that range. The only situation where
it would be necessary to use extension grade
wire instead of a external measuring junction is
where the junction box temperature is outside
the range of reference junction compensation
provided by the CR23X. This is only a factor
when using type K thermocouples, where the
upper limit of the reference compensation
linearization is 100°C and the upper limit of the
extension grade wire is 200°C. With the other
types of thermocouples, the reference
compensation range equals or is greater than
the extension wire range. In any case, errors
can arise if temperature gradients exist within
the junction box.
Figure 13.4-3 illustrates a typical junction box.
Terminal strips will be a different metal than the
thermocouple wire. Thus, if a temperature
gradient exists between A and A' or B and B',
the junction box will act as another
thermocouple in series, creating an error in the
voltage measured by the CR23X. This
thermoelectric offset voltage is a factor whether
or not the junction box is used for the reference.
It can be minimized by making the thermal
conduction between the two points large and
the distance small. The best solution in the
case where extension grade wire is being
connected to thermocouple wire would be to
use connectors which clamped the two wires in
contact with each other.
Summary of Contents for CR23X
Page 8: ...CR23X TABLE OF CONTENTS vi This is a blank page ...
Page 12: ...CR23X MICROLOGGER OVERVIEW OV 2 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE OV1 1 CR23X Micrologger ...
Page 34: ...CR23X MICROLOGGER OVERVIEW OV 24 This is a blank page ...
Page 50: ...SECTION 1 FUNCTIONAL MODES 1 16 This is a blank page ...
Page 72: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 8 This is a blank page ...
Page 88: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 10 This is a blank page ...
Page 110: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 22 This is a blank page ...
Page 134: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 24 This is a blank page ...
Page 164: ...SECTION 9 INPUT OUTPUT INSTRUCTIONS 9 30 This is a blankpage ...
Page 188: ...SECTION 11 OUTPUT PROCESSING INSTRUCTIONS 11 8 This is a blankp age ...
Page 221: ...SECTION 13 CR23X MEASUREMENTS 13 21 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Page 229: ...14 3 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE 14 3 1 CR23X Battery Pack and Panel ...
Page 240: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 14 This is a blank page ...
Page 244: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Page 268: ...APPENDIX B CONTROL PORT SERIAL I O INSTRUCTION 15 B 24 This is a blank page ...
Page 276: ...APPENDIX C BINARY TELECOMMUNICATIONS C 8 This is a blank page ...
Page 278: ...This is a blank page ...
Page 282: ...APPENDIX F DYNAGAGE SAP FLOW P67 F 4 This is a blank page ...
Page 299: ...APPENDIX I TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Page 300: ...This is a blank page ...
Page 302: ...This is a blank page ...
Page 308: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Page 324: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Page 340: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Page 342: ...This is a blank page ...
Page 373: ...This is a blank page ...