TD ADDENDUM—OVERVIEW
AD-OV-5
OV5.2 SAMPLE PROGRAM 2
This second example is more representative of a
real-life data collection situation. Once again the
internal temperature is measured, but it is used
as a reference temperature for the differential
voltage measurement of a type T (copper-
constantan) thermocouple; the CR10X should
have arrived with a short type T thermocouple
connected to differential channel 5.
When using a type T thermocouple, the copper
lead (blue) is connected to the high input of the
differential channel, and the constantan lead
(red) is connected to the low input.
A thermocouple produces a voltage that is
proportional to the difference in temperature
between the measurement and the reference
junctions.
To make a thermocouple (TC) temperature
measurement, the temperature of the reference
junction (in this example, the approximate panel
temperature) must be measured. The CR10X
takes the reference temperature, converts it to
the equivalent TC voltage relative to 0
°
C, adds
the measured TC voltage, and converts the
sum to temperature through a polynomial fit to
the TC output curve (Section 13.4).
The internal temperature of the CR10X is not a
suitable reference temperature for precision
thermocouple measurements
. It is used here
for the purpose of training only. To make
thermocouple measurements with the CR10X,
purchase the Campbell Scientific Thermocouple
Reference, Model CR10XCR (Section 13.4) and
make the reference temperature measurement
with Instruction 11.
Instruction 14 directs the CR10X to make a
differential TC temperature measurement. The
first parameter in Instruction 14 is the number of
times to repeat the measurement. Enter 1,
because in this example there is only one
thermocouple. If there were more than 1 TC,
they could be wired to sequential channels, and
the number of thermocouples entered for
repetitions. The CR10X would automatically
advance through the channels sequentially and
measure all of the thermocouples.
Parameter 2 is the voltage range to use when
making the measurement. The output of a type
T thermocouple is approximately 40 microvolts
per degree C difference in temperature between
the two junctions. The
±
2.5 mV scale will
provide a range of
±
2500/40 =
±
62.5
°
C (i.e.,
this scale will not overrange as long as the
measuring junction is within 62.5
°
C of the panel
temperature). The resolution of the
±
2.5 mV
range is 0.33 µV or 0.008
°
C.
Parameter 3 is the analog input channel on
which to make the first, and in this case only,
measurement. Parameter 4 is the code for the
type of thermocouple used. This information is
located on the Prompt Sheet or in the
description of Instruction 14 in Section 9. The
code for a type T (copper-constantan)
thermocouple is 1.
Parameter 5 is the Input Storage location in
which the reference temperature is stored.
Parameter 6 is the Input Storage location in
which to store the measurement (or the first
measurement; e.g., if there are 5 repetitions
and the first measurement is stored in location
3, the final measurement will be stored in
location 7). Parameters 7 and 8 are the
multiplier and offset. A multiplier of 1 and an
offset of 0 outputs the reading in degrees C. A
multiplier of 1.8 and an offset of 32 converts the
reading to degrees F.
In this example, the sensor is measured once a
minute, and the average temperature is output
every hour. Once a day the maximum and
minimum temperatures and the times they occur
will be output.
The first example described program entry one
keystroke at a time. This example does not
show the "A" key. Remember, "A" is used to
enter and/or advance (i.e., between each line in
the example below). This format is similar to
the format used in EDLOG.
It's a good idea to have both the manual and the
Prompt Sheet handy when going through this
example. You can find the program instructions
and parameters on the Prompt Sheet and can
read their complete definitions in the manual.
To obtain daily output, the Data Table
instruction is followed by the Output Instructions
to store the daily maximum and minimum
temperatures and the time each occurs.
Summary of Contents for CR23X
Page 8: ...CR23X TABLE OF CONTENTS vi This is a blank page ...
Page 12: ...CR23X MICROLOGGER OVERVIEW OV 2 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE OV1 1 CR23X Micrologger ...
Page 34: ...CR23X MICROLOGGER OVERVIEW OV 24 This is a blank page ...
Page 50: ...SECTION 1 FUNCTIONAL MODES 1 16 This is a blank page ...
Page 72: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 8 This is a blank page ...
Page 88: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 10 This is a blank page ...
Page 110: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 22 This is a blank page ...
Page 134: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 24 This is a blank page ...
Page 164: ...SECTION 9 INPUT OUTPUT INSTRUCTIONS 9 30 This is a blankpage ...
Page 188: ...SECTION 11 OUTPUT PROCESSING INSTRUCTIONS 11 8 This is a blankp age ...
Page 221: ...SECTION 13 CR23X MEASUREMENTS 13 21 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Page 229: ...14 3 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE 14 3 1 CR23X Battery Pack and Panel ...
Page 240: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 14 This is a blank page ...
Page 244: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Page 268: ...APPENDIX B CONTROL PORT SERIAL I O INSTRUCTION 15 B 24 This is a blank page ...
Page 276: ...APPENDIX C BINARY TELECOMMUNICATIONS C 8 This is a blank page ...
Page 278: ...This is a blank page ...
Page 282: ...APPENDIX F DYNAGAGE SAP FLOW P67 F 4 This is a blank page ...
Page 299: ...APPENDIX I TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Page 300: ...This is a blank page ...
Page 302: ...This is a blank page ...
Page 308: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Page 324: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Page 340: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Page 342: ...This is a blank page ...
Page 373: ...This is a blank page ...