SECTION 9. INPUT/OUTPUT INSTRUCTIONS
9-8
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
2
Repetitions
02:
2
Range code for
excitation
measurement (V
1
)
(Do not use autorange)
(Table 9-1)
03:
2
Range code for bridge
measurement (V
2
)
04:
2
Differential channel
number for first
measurement
05:
2
Excitation channel number
(Table 9-2A)
06:
4
Excitation voltage
(millivolts)
07:
4
Input location number
for first measurement
08:
FP
Multiplier
09:
FP
Offset
Input locations altered: 1 per repetition
*** 10 BATTERY VOLTAGE ***
FUNCTION
This instruction measures the CR23X input
voltage and writes it to an input location. The
units for voltage are volts. The measurement is
.2 to .3 volts less than actual battery voltage
measurement. See Section 14.2 for CR23X
power requirements.
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
4
Input location
Input locations altered: 1
*** 11 107 THERMISTOR PROBE ***
FUNCTION
This Instruction applies a 2.5 VAC excitation
voltage to Campbell Scientific's Model 107
Thermistor Probe, makes a slow single-ended
voltage measurement across a resistor in series
with the thermistor, and calculates the
temperature in °C with a polynomial. A
multiplier of 1 and an offset of zero yields
temperature in degrees Celsius. The maximum
polynomial error from -40°C to +56°C is given
here:
Curve Fit Error --
Range (°C)
Error (°C)
-40 to +56
<
±
1.0
-24 to +48
<
±
0.1
TABLE 9-3. Excitation/Integration Codes
Code Result
0x
excite all rep with channel x
1x
increment chan x with each rep
2x
excite all reps with channel x, slow 60 Hz
rejection, 10 ms delay
3x
excite all reps with channel x, slow 50 Hz
rejection, 10 ms delay
4x
increment chan x with each rep, slow 60 Hz
rejection, 10 ms delay
5x
increment chan x with each rep, slow 50 Hz
rejection, 10 ms delay
PARAM.
DATA
NUMBER
TYPE
DESCRIPTION
01:
2
Repetitions
02:
2
Single-ended channel
number of first
measurement
03:
2
Excitation channel/
Integration (Table 9-3)
04:
4
Input location for first
measurement
05:
FP
Multiplier
06:
FP
Offset
Input locations altered: 1 per repetition
*** 12 207 RELATIVE HUMIDITY PROBE ***
FUNCTION
This instruction applies a 3.0 VAC excitation
across Campbell Scientific's Model 207
Temperature and RH Probe, makes a fast
single-ended measurement across a series
resistor, calculates the result with a 5th order
polynomial, and performs the required
temperature compensation before outputting
the result in % RH.
When measuring several probes, all the RH
elements should be connected sequentially.
The temperature values used to correct the RH
measurements should also be stored
sequentially to make use of the REP feature in
Instruction 11.
Summary of Contents for CR23X
Page 8: ...CR23X TABLE OF CONTENTS vi This is a blank page ...
Page 12: ...CR23X MICROLOGGER OVERVIEW OV 2 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE OV1 1 CR23X Micrologger ...
Page 34: ...CR23X MICROLOGGER OVERVIEW OV 24 This is a blank page ...
Page 50: ...SECTION 1 FUNCTIONAL MODES 1 16 This is a blank page ...
Page 72: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 8 This is a blank page ...
Page 88: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 10 This is a blank page ...
Page 110: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 22 This is a blank page ...
Page 134: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 24 This is a blank page ...
Page 164: ...SECTION 9 INPUT OUTPUT INSTRUCTIONS 9 30 This is a blankpage ...
Page 188: ...SECTION 11 OUTPUT PROCESSING INSTRUCTIONS 11 8 This is a blankp age ...
Page 221: ...SECTION 13 CR23X MEASUREMENTS 13 21 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Page 229: ...14 3 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE 14 3 1 CR23X Battery Pack and Panel ...
Page 240: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 14 This is a blank page ...
Page 244: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Page 268: ...APPENDIX B CONTROL PORT SERIAL I O INSTRUCTION 15 B 24 This is a blank page ...
Page 276: ...APPENDIX C BINARY TELECOMMUNICATIONS C 8 This is a blank page ...
Page 278: ...This is a blank page ...
Page 282: ...APPENDIX F DYNAGAGE SAP FLOW P67 F 4 This is a blank page ...
Page 299: ...APPENDIX I TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Page 300: ...This is a blank page ...
Page 302: ...This is a blank page ...
Page 308: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Page 324: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Page 340: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Page 342: ...This is a blank page ...
Page 373: ...This is a blank page ...