SECTION 7. MEASUREMENT PROGRAMMING EXAMPLES
7-14
The following calculations are based on using a
Geokon model 4500 Vibrating Wire sensor. An
individual multiplier and offset must be
calculated for each sensor used in a system.
MULTIPLIER
The fundamental equation relating frequency to
pressure is
P = -F
x
G + B where
P = pressure, PSI
G = the Gage Factor obtained from the
sensors calibration sheet in PSI/digit.
The units of a digit are Hz
2
(10
-3
).
B = offset
F
x
= f
2
Hz
2
(10
-3
), where f is frequency.
Instruction 28 measures period, T, of the
vibrating wire in milliseconds (ms) and returns a
measured value, X, of
X = 1/(T
2
(ms)
2
) = f
2
(10
-6
)Hz
2
A multiplier of -1000 in Instruction 28 converts
the measurement to digits, as shown below.
-F
x
= -X(-10
3
) = -f
2
(10
-3
)Hz
2
To calculate the multiplier, convert Geokon's
gage factor, G, to the desired units (i.e., feet of
water per digit) and multiply by -1000 digits/kHz
2
.
TEMPERATURE CORRECTION
The temperature correction is applied as
follows.
P
T
= P + C
∗
(t
1
- t
0
), where
P
T
= Pressure corrected for temperature,
°
C
C
= Temperature coefficient, PSI/
°
C
(from Geokon calibration sheet)
t
0
& t
1
= Initial and current temperatures,
°
C.
The temperature coefficient, C, must be
converted to units compatible with the gage
factor, G.
WELL MONITORING EXAMPLE
In this example the vibrating wire sensor is used
to monitor water table height (Figure 7.15-2).
The desired data is the distance from the lip of
the well to the water surface. The sensor is
vented to atmosphere to eliminate
measurement errors due to changes in
barometric pressure. The water level is
expected to stay within 40 to 80 feet of the lip so
the 50 psi pressure sensor is placed
approximately 100 feet below the lip of the well.
The calibration data from Geokon is provided in
Table 7.15-1.
TABLE 7.15-1 Calibration Data for
Sensor 3998
Gage Factor
Temp. Coeff.
(psi/digit)
(psi/
°
C)
0.0151
-0.0698
The multiplier, m, is calculated to convert the
reading to feet of water.
m = 0.0151 (psi/digit)
∗
2.3067 (ft of water/psi)
∗
-1000 digits/kHz
2
= -34.831 ft of water/kHz
2
After the probe reaches thermal equilibrium, the
initial temperature, t
0
, is measured to be 24
°
C.
The water column above the sensor is referred
to as the "Reading". The Reading decreases
with increasing "Distance" from lip of well to
water surface so the Distance is computed by
subtracting the Reading from the Offset as
shown in Figure 7.15-2.
The "Initial Distance" to the water surface is
measured with a chalked line to be 47.23 feet
below the lip. The first time the program is
executed, the program calculates the offset
(Offset = Di Reading) required to obtain
a reading of 47.23 feet. The offset is stored in
Location 4 and applied to subsequent
measurements.
NOTE:
Following program compilation in
the
0
Mode, all input locations are set
to zero. This fact is utilized to detect the
first execution following a program
compilation.
The example assumes the sensor has been
connected as shown here.
Summary of Contents for CR23X
Page 8: ...CR23X TABLE OF CONTENTS vi This is a blank page ...
Page 12: ...CR23X MICROLOGGER OVERVIEW OV 2 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE OV1 1 CR23X Micrologger ...
Page 34: ...CR23X MICROLOGGER OVERVIEW OV 24 This is a blank page ...
Page 50: ...SECTION 1 FUNCTIONAL MODES 1 16 This is a blank page ...
Page 72: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 8 This is a blank page ...
Page 88: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 10 This is a blank page ...
Page 110: ...SECTION 7 MEASUREMENT PROGRAMMING EXAMPLES 7 22 This is a blank page ...
Page 134: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 24 This is a blank page ...
Page 164: ...SECTION 9 INPUT OUTPUT INSTRUCTIONS 9 30 This is a blankpage ...
Page 188: ...SECTION 11 OUTPUT PROCESSING INSTRUCTIONS 11 8 This is a blankp age ...
Page 221: ...SECTION 13 CR23X MEASUREMENTS 13 21 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Page 229: ...14 3 1 2 3 A 4 5 6 B 7 8 9 C 0 D FIGURE 14 3 1 CR23X Battery Pack and Panel ...
Page 240: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 14 This is a blank page ...
Page 244: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Page 268: ...APPENDIX B CONTROL PORT SERIAL I O INSTRUCTION 15 B 24 This is a blank page ...
Page 276: ...APPENDIX C BINARY TELECOMMUNICATIONS C 8 This is a blank page ...
Page 278: ...This is a blank page ...
Page 282: ...APPENDIX F DYNAGAGE SAP FLOW P67 F 4 This is a blank page ...
Page 299: ...APPENDIX I TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Page 300: ...This is a blank page ...
Page 302: ...This is a blank page ...
Page 308: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Page 324: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Page 340: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Page 342: ...This is a blank page ...
Page 373: ...This is a blank page ...