User Manual - Rev AJ
California Instruments
VOLT MAX
Standard Categories
A and Z
B 28V / 14V
RTCA 30.3 30.3 15.1
EUROCAE
30.3 29.3 14.6
Table 9-14: Normal Voltage Maximum
VOLT UNDER
This test applies to category Z and 28 volt category B equipment. The output voltage will drop to
10 volts and will ramp up at a rate of 0.15 volt/sec for the US standard and at a rate of 0.30
volt/sec for EUROCAE standard for 30 seconds before it returns to nominal value.
VOLT SURGE
This test will surge and sag the voltage to a
ation sp
5
seconds between transients. The test is repeated three times.
ategory
rge
Sags
This test will change the output voltage from 28V or 14V to 30.3V or 15.1V. The test will last for
30 minutes. The
←
(backspace) will terminate the test at any time.
level and dur
ecified Tab
with
C
Su
Volt
Dwell(msec)
Volt
Dwell(msec)
RTCA
EUR
S
EUR
U
A 40
40 30
5
17
30
1
B 40
40 30 15
17
30
Z 50
50 5
12
30
0 12
Table 9-15: Voltage Surge
POWER INTERRUPT
Refer to section POWER INTERRUPT.
9.3.1.6.2 Abnormal Test
From the DO-160 MENU scroll to the ABNORMAL DC entry using the up and down cursor keys.
Press the EN
y
ear as shown
TER ke to select the ABNORMAL screen. The screen will app
Figure 9-12: Abnormal State
The Abnormal Test has the following tests:
1. VOLT
MIN
2. VOLT
MAX
LT
LOW
4. VOLT
DROP
5. VOLT
SURGE
3. VO
MX Series
203