6.2 Front Panel Calibration Screens
The calibration screens for output or measurement calibration can be selected from the MENU 3
light
ems with three-phase output capability, use the PHASE key on the front panel to
ame
icated above.
screen. (Press MENU button several times to toggle to MENU 3 screen.)
To select the OUTPUT CALIBRATION screen press the
↑
or
↓
key several times to high
OUTPUT CAL. Then press the ENTER key. This will bring up the PASSWORD screen. To
prevent unauthorized access to calibration data, a password must be entered to access any
calibration screen. The calibration password is always "
5000
" and may be entered using the
numeric keypad. Once entered, the calibration screens remain accessible until the MX unit is
powered down.
Type 5000 and press the ENTER key to show the OUTPUT CALIBRATION screen.
On MX syst
select the phase to be calibrated.
To select the MEASUREMENT CALIBRATION screen, follow the same steps as outlined above
but select the MEASUREMENT CAL entry instead of OUTPUT CAL. If another CALIBRATION
screen has been accessed since power-up, no password is needed. Otherwise, enter the s
password as ind