CLOCK MODE
Selects internal or external clock source. The MX Series
controller uses an open-air crystal time base with an accuracy
of 100 ppm. To improve output frequency stability and
se
power ratings are divided by two.
DC OFFSET
When the AC+DC mode is selected, the VOLTAGE field in the
PROGRAM 1 screen is used to set the AC portion of the output
voltage. The DC OFFSET field in the PROGRAM 2 screen can
be used to set the DC offset level. Either the knob or the
decimal keypad may be used to set the DC offset level.
START ø
made to either
accuracy, an external clock generator may be used.
VOLT MODE
The MX Series offers three output modes, AC, DC and AC+DC.
The VOLT MODE field can be used to toggle between the
three output modes. Both the Knob and the +/- key may be
used to toggle through these three selections. In DC mode, no
frequency selection is possible and all maximum current and
Selects the start phase angle for output changes
voltage or frequency. This allows changing the output at a
specific phase angle. The output on key also uses this phase
angle setting to program the output voltage up to the set level
after the output relay is closed. The default value for this field is
RANDOM.