Output Parameter
MX30
MX45
MX60
MX90
MX135
Resolution:
0.01 Hz
from 16.00 to 81.91 Hz
0.1 Hz
from 82.0 to 819.0 Hz
Accuracy:
± 0.01 %
Phase (3 phase mode)
Range:
Phase B/C relative to phase A
0.0 to 360.0°
Resolution: 0.1°
Accuracy:
16 Hz - 100 Hz: < 1.5°
100 Hz - 500 Hz: < 2°
> 500 Hz:
< 4°
Ext. Sync Mode
Input:
Isolated TTL input for external frequency control. Requires 5V at 5 mA for logic
high.
Accuracy:
Ext. Sync to phase A with fixed Ext. Sync Frequency input:
16 Hz - 100 Hz: < 2°
100 Hz - 500 Hz: < 3°
> 500 Hz:
< 4°
Programmable Output Impedance
(available on MX30-3Pi and MX45-3Pi in three phase mode only)
Range:
R: 1 – 200 mOhm
L: 15 – 200 uH
N/A N/A N/A
Resolution:
R: 1 mOhm
L: 1 uH
N/A N/A N/A
Accuracy:
10 % FS
N/A
N/A
N/A
Note: Output specifications apply below the Current / Voltage rating lines shown in the
V/I rating chart below.
Figure 2-1: MX30-3 Voltage / Current Rating Chart for 150/300 V AC Ranges.
16 MX
Series