4.2.14 OUTPUT CAL FACTORS Menu
Figure 4-29: OUTPUT CAL FACTORS Menu (Series I only)
Figure 4-30: OUTPUT CAL FACTORS Menu (Series II only)
The OUTPUT CAL FACTORS menu provides access to the output calibration parameters.
These parameters are password protected and can only be changed after the calibration
password has been entered. For three phase configurations, the PHASE keys toggle between
the three calibration screens for each phase. Refer to the calibration section in this manual for
details on performing a calibration.
The following calibration factors are available from this menu:
Entry Description
IHARM FS
Option –413 Inter harmonic calibration coefficient. (Series II
models only.)
IMP. REAL FS
Full scale resistive output impedance calibration factor.
IMP. REACT FS
Full scale inductive output impedance calibration factor.
IMP. REAL MIN
Minimum resistive AC source output impedance. The AC
source has an output impedance greater than zero. This value
determines the minimum resistive component of the AC source
output impedance.
IMP. REACT M
ance. The AC source
s
VOLT FS
Full scale voltage output calibration factor.
V HI-FREQ
High frequency full-scale voltage output calibration factor.
(Series I models only.)
VOLT ZERO
Zero offset voltage calibration factor.
PHASE OFST
Phase offset calibration factor. Compensates for phase shift
caused by AC amplifier.
IN
Minimum reactive AC source output imped
has output impedance greater than zero. This value determine
the minimum inductive component of the AC source output
impedance.