This test will set the voltage at 115V and will remain at this voltage through out the test. The
9.3.1.6
test will cycle the frequency three times as shown in Table 9-12. Each repeat will start from
sequence 2.
DC TESTS
If the output voltage is set for 24V DC or 14V DC the DO-160 DC Main selection screen will
appear as seen in Figure 9-10.
Figure 9-10: DO-160 DC Main Menu
Standard selection and Category selection are required. Use
TCA or EUROCAE. Also, select equipment category A, B or Z.
9.3.1.6.1 Normal State Test
up and down cursor keys. Press the ENTER key
reen will appear as shown in Figure 9-11.
Note: Prior to test selection the
the shuttle to select Standard R
Scroll to the NORMAL STATE entry using the
to select the NORMAL STATE screen. The sc
Figure 9-11: Normal State
AL screen has the following tests:
2. VOLT
MAX
3. VOLT
UNDER
4. VOLT
SURGE
5. POWER
INTERRUPT
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test. For some of these tests, numeric data entry
may be required to define the test number or the modulation rate.
VOLT MIN
The DO-160 NORM
1. VOLT
MIN
Standard
Categories
A and Z
B 28V / 14V
RTCA 22.0 22.0 11.0
EUROCAE
22.0 25.1 12.5
Table 9-13: Normal Voltage Minimum
This test will change the output voltage from 28V or 14V to 22V or 11V. The test will last for 30
minutes. The
←
(backspace) will terminate the test at any time.