User Manual – Rev AJ
California Instruments
4.2.12 OUTPUT IMPEDANCE Menu [
MX30-3Pi or MX45-3Pi Model in 3 phase mode only
]
Figure 4-26: OUTPUT IMPEDANCE Menu
MX45-3Pi offers programmable output impedance in three-phase mode of
ns by programming resistive
nce function is available in single phase mode or on MX
The MX30-3Pi and
operation. This allows the user to simulate line impedance conditio
and inductive elements of the AC source’s output impedance.
Note: No programmable impeda
system configurations other than the MX35-3Pi.
The OUTPUT IMPEDANCE menu can be accessed from MENU 3. The following fields are
available in this menu:
Entry Description
ance.
(The lower limits for the resistive and inductive components
may vary on different AC source models and are set as part of
the source calibration.)
INDUCTIVE
Sets the inductive component of the output impedance.
(The lower limits for the resistive and inductive components
may vary on different AC source models and are set as part of
the source calibration.)
SET MINIMUM
Selecting this field and pressing the ENTER key sets both
omponents to their minimum values. This
ns the programmable impedance function off.
function of the MX30-3Pi or MX45-3Pi is mutually
RESISTIVE
Sets the resistive component of the output imped
impedance c
effectively tur
Note: The Programmable Impedance
exclusive with the Automatic Voltage Level Control (ALC) mode of operation. One
nable the other.
must be turned off the e
116 MX
Series