![Cadex C8000 User Manual Download Page 120](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378120.webp)
C8000 Battery Testing System User Manual
108
3.4
Test Batteries with BatteryLab
To test a battery:
1.
From the
BatteryLab Main
menu, click
Connect
.
The
C8000
window appears.
BatteryLab waits for C8000s to connect. When a C8000 connects to
BatteryLab, the C8000 window title updates to reflect the IP
Address of the connected C8000. The C8000 also updates its Title
Header from
C8000
to the
Unit #
indicated on the analyzer
window. Wait for the C8000 to connect to BatteryLab before
starting to test a battery.
If the C8000 and software do not connect within 15 to 60 seconds,
check the IP addresses and verify that no IP address is duplicated
on the network. (An error message appears if the time-out limit is
reached.) Recycle power on the C8000, and restart BatteryLab. If
problems persist, see
1.
Select a C8000, choose a channel, select and right-click on
the channel and choose
Test Battery
. The
Test Battery
window appears.
Summary of Contents for C8000
Page 12: ......