![Cadex C8000 User Manual Download Page 44](http://html1.mh-extra.com/html/cadex/c8000/c8000_user-manual_3250378044.webp)
C8000 Battery Testing System User Manual
32
4.
If the filament inside the glass tube is broken, replace the
fuse. Be aware that the filament can sometimes appear
intact even when it is broken.
5.
With the fuse inserted, push the fuse cover back into place.
6.
Connect the power cord and turn ON the C8000.
2.6.4
Replace the Backup Battery
Caution!
Test results may be lost when the backup battery is
removed. Backup the calibration data to the PC hard
disk. Retrieve test results stored on the C8000 to the
BatteryLab database. To download the test results to
BatteryLab, select
Retrieve Results
from the
File
menu
on the
Test Results
window. For more information,
3.6 View Test Results in BatteryLab
The C8000 uses a 3-volt lithium coin battery (type CR2032 or
equivalent, Cadex part number 45-206-0001) to preserve settings
while the C8000 is turned OFF. With normal use of the C8000,
you will seldom need to replace the backup battery.
You will need the following items to replace the battery:
A #2 Phillips screwdriver.
A new 3-volt lithium battery.
A wrist strap or other device to eliminate electrostatic discharge.
Summary of Contents for C8000
Page 12: ......